Efficacy of water filters for dental chair units: Assessment of the filtration action versus Coxsackievirus B5
Articolo
Data di Pubblicazione:
2020
Abstract:
Introduction. The microbiological safety and control of the water used in dental practice has a critical importance for avoiding crosslinked infections in the dental office. The aim of this study was to establish coxsackie virus filtration of the water applied to a dental unit. Methods. A specific water filter-system was used, to verify the viral load in the outgoing water. The statistical analysis was performedusing the Shapiro-Wilk and t-Student test. Results. The outcome of the evaluation of the virologic tests shows an excellent capability of virus filtration that attested 99.9999% in the volume analyzed. A statistical difference was found in the bacterial water contamination parameter before and after filtration. (P!=!0.000000). Conclusions. According to the tests, medical devices applied to a dental unit are able to filter viruses and therefore reduce risk of contamination in the dental office.
Tipologia CRIS:
1.1 Articolo in rivista
Keywords:
Biological safety; Coxsackie virus; Dental chair; Hydraulic system; Microbiological tests
Elenco autori:
Scarano, A.; Murmura, G.; Lorusso, F.; Di Carlo, T.; Oliva, S.
Link alla scheda completa:
Link al Full Text:
Pubblicato in: